70L / Min MR340E Fisher Paykel Humidification Chamber For Infant Pediatric

Ask Latest Price
Video Channel
Brand Name :Fisher& Paykel
Model Number :MR340E
MOQ :1pcs
Payment Terms :T/T(wire transfer), Western Union, Paypal
Delivery Time :2-3 working days
Packaging Details :Carton
Product Name :MR340E Fisher& Paykel Humidification Chamber For Infant/Pediatric
Brand :Fisher& Paykel
Part No :MR340E
Condition :New
Maximum Operating :20kPa
Compliance :0.15-0.35mL/cm H2O
Maximum Peak Flow :70L/min
Maximum water capacity with scroll :230mL
Maximum water capacity without scroll :235mL
Warranty :90 days
Delivery Method :DHL,UPS ,Fedex ,etc
MOQ :One piece
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

MR340E Fisher& Paykel Humidification Chamber For Infant/Pediatric

MR340 is a reusable humidification chamber for use with Fisher & Paykel heated bases. This chamber is used to ensure the humidification of intubated or non-intubated patients. Its low compressible volume makes the MR340 ideal for infants or newborn applications.

Product Specification

Gas ports Inlet 15mm Female
Outlet 22mm Male
Maximum Operating 20kPa
Compliance 0.15-0.35mL/cm H2O
Maximum Peak Flow 70L/min

With Scroll Without scroll

Compressible Volume

Full 130mL 135mL
Empty 360mL 370mL
Maximum water capacity 230mL 235mL

Material Used In Construction

Base Aluminium
Scroll Aluminium
Wick Absorbent Paper
Dome PolysulPHiliphhilipone
Gasket Synthetic Rubber

Test Condition

Humidifier :MR730(similar results can be expected using an MR850/MR700/720 )

Humidification chamber :MR340 without scroll

Maximum Heater Plate Setting :110 ℃ using 150w heater plate

Ambient air temperature :22-24℃

With Heater wire

Recommended flowrates (L/min)

1.5 L/mim

(Minimum)

30L/min

(typical )

70L/min

(Maximum)

Pressure Drop(cm H2O) 0.01 0.38 1.96

Time to Rearch Airway

Temperature =37℃

5 min 5 min 5 min

Humidity at airway outlet

Set Temp=37℃

44mg/L 43.5mg/L 42mg/L

Without Heater wire

Recommended flowrates (L/min)

5 L/mim

(Minimum)

30L/min

(typical )

50L/min

(Maximum)

Pressure Drop(cm H2O) 0.01 0.38 1.02

Time to Rearch Airway

Temperature =37℃

<20 min 15 min 15 min

Humidity at airway outlet

Set Temp=37℃

44mg/L(37°) 43.5mg/L(36°) 42mg/L(35°)

Humidifier :MR410

Humidification chamber :MR340 with scroll

Ambient air temperature :22-24℃

Circuit:1.3*15mm Smoothbore Tubing (900MR529)

Recommended flowrates (L/min)

5 L/mim

(Minimum)

20L/min

(typical )

40L/min

(Maximum)

Pressure Drop(cm H2O) 0.01 0.05 0.18
Time to Reach Operating Temperature 25 min 30 min 40 min

Liquid Output(mg/L)

(Airway Temp Reached ℃)

Power Setting=5

Power setting =9

30mg/L(29.5°)

39mg/L(35°)

43.5mg/L(32.5°)

46mg/L(38°)

42mg/L(30°)

34mg/L(32°)

Warning

  • Do not exceed the maximum fill level. If the chamber is overfilled, fluid can enter the breathing circuit.
  • Do not fill the room with water above 37°C.
  • If using a water supply unit, make sure to place the water bag below the level of the humidifier after filling the water chamber.
  • Make sure to regularly monitor the water level in the chamber. Supplement if necessary.
  • Make sure the humidifier is not tilted. Tilting the humidifier may cause water to enter the breathing circuit.
  • Do not touch the heater plate or the chamber base. Surfaces may exceed 85°C.
  • Do not exceed the maximum peak flow of 70L/min.

Cleaning

The cleaning chamber can be cleaned by any of the following methods

Autoclave: 136°C (277°F) @ 220kPa (32psi) for 4 minutes or 120°C (248°F) @ 96kPa (14.1psi) for 15 minutes

Ethylene oxide: 55°C (131°F)

Pasteurization Chemical Immersion:

Solutions such as Mesoxazine, Pyrrole, Control 3, Solution 2 and Cidex have been tested according to the manufacturer's instructions. Other brand names with similar active ingredients can also be substituted.