China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group Win-win together!

Manufacturer from China
Verified Supplier
5 Years
Home / Products / Defibrillator Machine Parts /

Black External Paddles For Defibrillator Innomed Mod Cardio-Aid-200B

Contact Now
China World Technology Medical Equipment Service Group
Visit Website
City:guangzhou
Province/State:guangdong
Country/Region:china
Contact Person:MissKiara
Contact Now

Black External Paddles For Defibrillator Innomed Mod Cardio-Aid-200B

Ask Latest Price
Video Channel
Brand Name :Innomed
Model Number :Cardio-Aid-200B
MOQ :1pcs
Payment Terms :T/T(wire transfer), Western Union, Paypal
Delivery Time :3-5 working days
Packaging Details :Carton
Product Name :External Paddles For Defibrillator Innomed, Mod. Cardio-Aid-200B
Brand :Innomed
Model No. :Cardio-Aid-200B
MOQ :One piece
Warranty :90 days
Condition :Used
Delivery Method :DHL,UPS ,Fedex ,etc
Payment Method :T/T ,Paypal ,Western Union
Packaging :Carton
Color :Black
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

External Paddles For Defibrillator Innomed, Mod. Cardio-Aid-200B

The right blade (Apex) contains the charge and shock key. The charging key is labeled "2" and is called "Key 2" on the screen. The shock key is marked "3" and mentioned again on the screen.

The left (sternal) plate contains a shock key labeled "3". There are also plus and minus buttons labeled "Power" and these buttons only work in manual mode. These buttons will increase or decrease the power level and the actual value can be shown on the display. If either is pressed in manual mode, the device will deactivate while charging or preparing for a discharge.

Functional test of the paddle version

1. Make sure the device is connected to the electrical network.

2. Turn on the defibrillator.

3. Connect the electrodes to a known ECG test signal.

4. Verify that the ECG bar is displayed.

5. Verify that the heart rate on the screen matches the heart rate on the simulator.

6. Connect the ECG cable to the ECG input.

7. Connect the ECG cables to a known ECG test signal.

8. Verify that the ECG bar is displayed.

9. Check that the heart rate on the screen is the same as the heart rate on the simulator.

10. Activate the "Print" button to check the ECG test signal printed by the recorder.

11. Make sure the paddle is inserted into the paddle holder.

12. Set the mode to Manual and select 360J with the blade.

13. Press the load button and wait for the "ready" signal.

14. Press both "Download" buttons at the same time.

Grades! The integrated test system compares the selected energy with the actual energy measured during discharge. The result appears on the screen and the machine prints the receipt. If discrepancies are found, an error message is displayed.

Item name

external paddles for defibrillator Innomed, mod. Cardio-Aid-200B

Brand Innomed
Model Cardio-Aid-200B
Use place Hospital Devices; Medical Care; ICU;
Deliver time 3-5 days
Warranty 90 days

External Paddles test.

Discharge in open air

1. Hold the paddle in the air.

2. Select and load 5J

3. Download attempt and impossible confirmation screen: "No patient contact"

4. Press Disable on Cardio-Aid®.

Defi tester

Delivers a 50J shock to the onboard defibrillator. The delivered energy is measured, displayed on the screen and printed out (if selected in the configuration menu). Check that the delivered energy corresponds to the selected energy and that the test result shows PASS. Approximate tolerances are listed in the table below.

Sound check

1.Put the paddle on QA 45

2. Select 2J and charge

3. Verify that the tone explodes during charging and ends with a steady tone when the defibrillator is ready

Defibrillator safety

Note: Make sure both paddles are connected to the built-in defi tester or external power meter (50 ohm load).

1.Select 2J and charge

2. Press one push button and then the other (one at a time) and confirm no power is supplied. Tap Disable.

Accuracy of delivered energy

Select the specified energy level and download the energy meter. Check that the supplied current is within the tolerance limits.

NOTE: The energy shock should be administered immediately after the defibrillator has entered standby mode.

Defi Ready got timeout.

1. Charge to 5J.

2. Check if the energy is available in 30s. +/- 2 sec

3. Check if Cardio-Aid® automatically deactivates.

Disarm.

1. Charge to 360J

2. When Ready appears, tap Disable

3. Confirm that "Disconnect" is displayed and a loading sequence can be restarted.

max. charging time on the power pack.

1. Charge to 360J

2. Check that the charging time does not exceed 7 seconds.

3. Discharge via QA 45

max. battery charging time (@ 25deg.C).

1. Unplug the power cord

2. Charge to 360J

3. Check that the charging time does not exceed 7 seconds. (usually) with a full

charged battery.

4. Discharge via QA 45

If you're interested in External Paddles For Defibrillator Innomed, Mod. Cardio-Aid-200B,feel free to contact us .Thank you .