China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group Win-win together!

Manufacturer from China
Verified Supplier
5 Years
Home / Products / Patient Monitor Accessories /

2016998-001 GE Dual Reusable Temperature Adapter Cables For 400 Series 50cm 20 Inch

Contact Now
China World Technology Medical Equipment Service Group
Visit Website
City:guangzhou
Province/State:guangdong
Country/Region:china
Contact Person:MissKiara
Contact Now

2016998-001 GE Dual Reusable Temperature Adapter Cables For 400 Series 50cm 20 Inch

Ask Latest Price
Video Channel
Brand Name :GE
Model Number :2016998-001
MOQ :1pcs
Payment Terms :Full before shipment
Delivery Time :3-5 working days
Packaging Details :Carton
Product Name :GE
Part No :2016998-001
Use for :400 Series
Probe length :50cm(20 inch)
Condition :New
Warranty :90 days
MOQ :One piece
Payment Method :T/T ,Paypal , western union
Delivery Method :DHL,UPS ,Fedex ,etc
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

2016998-001 GE Dual Reusable Temperature Adapter Cables For 400 Series ,50cm(20 Inch)

Two temperatures are measured by two functions. It ensures the alarm of the patient with adjustable high and low temperature limits. There are system alarms for sensor and calibration errors.

The temperature connector and measurement circuit are used in conjunction with the cardiac output monitor function. So you cannot use both functions at the same time. The signal in the patient cable indicates proper operation.

Temperature calibration

NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from GE Healthcare, order code 884515-HEL. A Dual temperature adapter cable, order code 2016998-001 is also required for the temperature calibration. Calibrate the temperature, when the measured test values differ for more than ±0.1 °C, and always after STP board replacement.

1. Enter the STP service menu. (Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) - Parameters).

2. Enter Calibrations menu.

  • Choose Protection OFF in protect mode.
  • Select Calibrate T1/Calibrate T2.
  • Insert calibration plug (25 °C) into T1/T2 connector.
  • Push the ComWheel.
  • Insert calibration plug (45 °C) into T1/T2 connector.
  • Push the ComWheel.

9. Choose Protection ON in protect mode

GE Dual Reusable Temperature Adapter cables for 400 series ,50cm(20 inch) 2016998-001 specification

Product Name GE Dual Reusable Temperature Adapter cables for 400 series ,50cm(20 inch) 2016998-001
Brand GE
Part No 2016998-001
Use for 400 Series
Probe length 50cm(20 inch)

Temperature test producer

1. Set the temperature output of the patient simulator to 37° C.

2. Connect the temperature adapter cable to the TEMP/CO connector of the patient monitor.

3. Set the switch on the temperature adapter to the 400 or 700 position according to customer preference.

4. Connect the 400 or 700 series temperature output connector of the patient simulator to the T1 connector of the temperature adapter.

5. Check that the TEMP parameter window appears on the patient monitor display with a T1 reading of 37.0° ± 0.4° C.

6. Move the temperature simulator cable from the T1 connector of the temperature adapter to the T2 connector of the temperature adapter.

7. Check that the T2 reading in the TEMP parameter window on the patient monitor display is 37.0°±0.4°C

Two temperatures are measured by two functions. It ensures the alarm of the patient with adjustable high and low temperature limits. There are system alarms for sensor and calibration errors.

The temperature connector and measurement circuit are used in conjunction with the cardiac output monitor function. So you cannot use both functions at the same time. The signal in the patient cable indicates proper operation.

Temperature calibration

NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from GE Healthcare, order code 884515-HEL. A Dual temperature adapter cable, order code 2016998-001 is also required for the temperature calibration. Calibrate the temperature, when the measured test values differ for more than ±0.1 °C, and always after STP board replacement.

1. Enter the STP service menu. (Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) - Parameters).

2. Enter Calibrations menu.

  • Choose Protection OFF in protect mode.
  • Select Calibrate T1/Calibrate T2.
  • Insert calibration plug (25 °C) into T1/T2 connector.
  • Push the ComWheel.
  • Insert calibration plug (45 °C) into T1/T2 connector.
  • Push the ComWheel.

9. Choose Protection ON in protect mode

GE Dual Reusable Temperature Adapter cables for 400 series ,50cm(20 inch) 2016998-001 specification

Product Name GE Dual Reusable Temperature Adapter cables for 400 series ,50cm(20 inch) 2016998-001
Brand GE
Part No 2016998-001
Use for 400 Series
Probe length 50cm(20 inch)

2016998-001 GE Dual Reusable Temperature Adapter Cables For 400 Series 50cm 20 Inch

Temperature test producer

1. Set the temperature output of the patient simulator to 37° C.

2. Connect the temperature adapter cable to the TEMP/CO connector of the patient monitor.

3. Set the switch on the temperature adapter to the 400 or 700 position according to customer preference.

4. Connect the 400 or 700 series temperature output connector of the patient simulator to the T1 connector of the temperature adapter.

5. Check that the TEMP parameter window appears on the patient monitor display with a T1 reading of 37.0° ± 0.4° C.

6. Move the temperature simulator cable from the T1 connector of the temperature adapter to the T2 connector of the temperature adapter.

7. Check that the T2 reading in the TEMP parameter window on the patient monitor display is 37.0°±0.4°C

2016998-001 GE Dual Reusable Temperature Adapter Cables For 400 Series 50cm 20 Inch

8.Remove the temperature adapter and temperature simulator cable from the patient monitor and patient simulator.