Mindary BeneVision N17 Patient Monitor Parts Power Switch Button Board 050-002302-00

Ask Latest Price
Video Channel
Brand Name :Mindray
Model Number :050-002302-00
MOQ :1pcs
Payment Terms :Full before shipment
Delivery Time :3-5 working days
Packaging Details :Carton
Product Name :050-002302-00 Mindary BeneVision N17 Patient Monitor Power Switch Button and Power Switch Button Board
Brand :Mindray
Part No . :050-002302-00
Use for :Mindary BeneVision N17 Patient Monitor
Condition :Work in Good
Warranty :90 days
MOQ :One piece
Packaging :Standard
Delivery Method :DHL,UPS ,Fedex
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

050-002302-00 Mindary BeneVision N17 Patient Monitor Power Switch Button and Power Switch Button Board

Power Switch Board

There are the power switch and three indicators on the power switch board, which are the AC on-line indicator, battery indicator, and the power-on indicator

Power switch

◆ Pressing this switch turns on the monitor.

◆ When the monitor is on, pressing and holding this switch turns off the monitor.

Power On/Off Failures

Failure Symptom Possible Cause Countermeasure
Power on failure AC mains not connected or insufficient battery power or battery damaged

n Verify that AC power is properly connected.

Check whether the battery capacity is sufficient and whether the battery is damaged.

Cable defective or improperly connected

nCheck the cable connecting the power on/off button board to the main control board.

Verify that the connectors and corresponding sockets connecting the cables are not damaged.

Power switch and indicator board damaged Replace the power switch and indicator board.
Power module defective Replace the power module
Main control board failure Replace the main control board

Removing the Power Switch Board

1. Remove the cable on the power switch board.

2. Remove the two PT3×8 screws and take out the power switch board

Mindary BeneVision N17 Patient Monitor Parts Power Switch Button Board 050-002302-00

Note

  • Be careful when disconnecting the cable to avoid damage. When installing, align the positioning ribs of the front shell plate and
  • press the plate towards the direction of the silicone button. l
  • When reinstalling the board, be careful to avoid overtightening the screws and stripping the holes.

Care and Cleaning Safety Information

Warning

• Never immerse any part of the device or its accessories in liquid or allow liquid to get inside.

• Contact of cleaning agents or disinfectants with connectors or metal parts can cause corrosion.

• Never pour or spray any liquid directly on the device or accessories, or allow liquid to enter any connection or opening.

• If liquid is spilled on the device or accessories, unplug it, dry it, and contact service.

• Never use abrasive materials (such as steel wool or silver polish) or aggressive cleaning agents .

• Dilute and use cleaning agents or disinfectants according to the manufacturer's instructions.

• Check equipment after cleaning and disinfection. Discontinue use if there are any signs of damage.