Display LTM08C351S for Nihon Kohden BSM-2301C
Nihon Kohden BSM-2301C Life Scope Bedside Patient Monitor Display Assembly LTM08C351S Used-Good Condition
► Brief Description
Nihon Kohden BSM-2301C Life Scope Bedside Patient Monitor Display Assembly LTM08C351S is the most important part of the device, which can clearly display all kinds of detected data and waveforms. It helps medical staff to view the patient's condition more intuitively and conveniently. And the display can present settings such as operation menus to facilitate staff operations.
► Specification
Size: | 8.4 inch, TFT type color LCD |
Viewing area: | 170.4 mm × 127.8 mm |
Waveform display mode: | Non-fade moving or non-fade fixed |
Resolution: | 800 × 600 dots |
Maximum number of waveform trace: | 5 traces |
Sweep speed: | 25 mm/s, 50 mm/s (Respiration and CO2 low speed: 1.56 mm/s, 6.25 mm/s) |
Sweep width: | about 124 mm at 25 mm/s sweep speed |
Waveform display color: | 12 |
Numeric display color: | 12 |
Waveform freeze: | Provided |
Display waveforms: | ECG, respiration, IBP, SpO2 pulse wave and CO2 |
Display sensitivity: | 10 mm/mV ±5% (at ×1 sensitivity) |
Sensitivity control: | ×1/4, ×1/2, ×1, ×2, ×4, or AUTO |
Pacing spike display: | Available |
Display Temperature range: | 0°C to 45°C (32 to 113°F) |
Display update cycle: | Every 3 s or when alarm is generated |
► Detailed Photos
► Display Problems
During the use of the Nihon Kohden BSM-2301C Life Scope Bedside Patient Monitor Display Assembly LTM08C351S, some problems may occur. Here are some possible problems and solutions:
- Wrong connection between CN0802 on the main board and the inverter. Check the connection.
- Wrong connection between the inverter and the LCD unit. Check the connection.
- Faulty inverter board. Replace the inverter board.
- Faulty backlight. Replace the backlight.
- Faulty connection between CN1001 on the main board and the LCD unit. Check the connection.
- Faulty LCD unit. Replace the LCD unit.
- Faulty main board. Replace the main board.
- Faulty main board. Replace the main board.
- Battery powered. Set Power Save Mode to Off on the System Settings screen.
- There is dirt on the touch screen (internal or external). Clean the touch screen surface.
- The backlight is near the end of its life (50,000 hours of life). Replace the backlight.
- There is a connection error between CN0802 on the main board and the inverter board. Just check the connection.
- Main board failure. Replace the main board.
- Analog board failure. Replace the analog board.