► Brief Introduction of Drager 8410580 Expiratory Valve New Original for Evita XL Evita XL Neo Ventilator
The Drager Exhalation Valve 8410580 is a original new part designed for the Evita XL and Evita XL Neo ventilators. This valve is essential for regulating the flow of exhaled air from the patient's lungs during mechanical ventilation.
Replacing your exhalation valve with this new genuine part will ensure optimal performance and efficiency of your Drager ventilation system. These high-quality components are manufactured to original equipment manufacturer standards and specifications.
It is recommended that the exhalation valve be checked and replaced regularly as part of routine maintenance to ensure proper operation of the ventilator and to minimize the risk of device failure.
►Specification of Drager 8410580 Expiratory Valve New Original for Evita XL Evita XL Neo Ventilator
Product Name | Drager 8410580 Expiratory Valve New Original for Evita XL Evita XL Neo Ventilator
|
Brand | Drager |
Part No | 8410580 |
Compatible with | Evita XL Evita XL Neo Ventilator
|
Reusable or not | Reusable |
MOQ | One piece |
Condition | Original new |
Warranty | 90 days |
Material | Plastic |
Color | White |
► Removing expiratory valve
1 Push the catch (A) to the right while
2 at the same time pulling out the expiratory valve (B).
CAUTION Only the white reusable expiratory valve may be reprocessed. Dispose of the blue disposable expiratory valve as infectious special waste.
3 If using the reusable expiratory valve, remove the optional collection container (C) from the water trap.
► Installing the expiratory valve
1. Tilt the control panel upward by pressing the left and right segments (A).
2. Push the exhalation valve (B) into the bracket until it clicks into place.
Note : the expiratory valve can only be inserted or removed when the flap is open. Keep the flap closed during ventilation.
► Sterilizing reusable expiratory valve
Once dry, sterilize the reusable exhalation valve in hot steam (134 °C (273 °F) for at least 10 minutes). Otherwise, liquid may remain in the pressure measuring line and affect its function.
Place the opened exhalation valve in a tray to prevent it from being damaged by other components.